Evaluation of Microwave
 
To evaluate your microwave please perform the following experiment:
  1. Place 40µl of DI water in the same size and type of tube as for PCR.
  2. Overlay each tube with mineral oil
  3. Close the tubes, place in microwave safe rack and heat on HIGH for 5 minutes.
  4. If any caps pop or tubes distort in any manner, then place a separate beaker in the microwave with 150ml of ambient temperature DI water and repeat as above. The beaker of water serves as a heat ballast.
  5. If tubes open or distort after performing the step above, then reduce power by 10% increments and increase time by 1 minute increments repeating the step above until tubes no longer open or distort.
If a water ballast was used as above, then allow the tubes to remain in the microwave undisturbed for 2 minutes following the completion of the microwave cycles. This is to protect you from any sudden boiling of the ballast after it has been heated.